TYPE |
Insulation resistance meter |
PHYSICAL SPECIFICATIONS |
19”W x 7”H x 15”D, (48 cm x 17 cm x 38 cm); Weight: 24lbs (10.9 kg) |
INPUT POWER |
110 – 120 Vac or 220 – 240 Vac (selectable), 50/60Hz |
RESISTANCE READING RANGE (0°C to +30°C) |
100 K-ohms – 1 M-ohm ±20%, 1 M-ohm – 1 T-ohm ±5%, 1 T-ohm – 5 T-ohm ±20% , 5kV 100 K-ohms – 1 M-ohm ±20%, 1 M-ohm – 100 G-ohms ±5%, 100 G-ohms – 500 G-ohms ±5%, 500 V 1 M-ohm – 10 G-ohms ±5%, 50 V |
TEST VOLTAGE |
Selectable from 50Vdc – 5 KVdc, in 2 Vdc steps |
OUTPUT VOLTAGE ACCURACY (0°C to +30°C) |
±2%, ±1v of selected voltage with load resistance greater than 100 MΩ |
SHORT CIRCUIT CURRENT |
2 mA max |
CURRENT READING RANGE (0°C to +30°C) |
0.03 nA – 2 mA; Accuracy: ±5% ±0.2 nA |
CAPACITANCE READING RANGE (0°C to +30°C) |
0.01 μF – 10.0 μF (Test voltage greater than 100V); Accuracy: ±5% ±0.03 μF |
VOLTAGE READING RANGE (0°C to +30°C) |
50 – 1250 Vac (rms) or dc; Accuracy: ±5%, ±1V |
CAPACITOR DISCHARGE |
Less than 2 Sec/μF, automatic at the end of test |
HUM REJECTION |
1mA per 1kv of test voltage, 2mA rms maximum |
BATTERIES |
Two 12V, 2.0 Ah sealed lead acid batteries. Battery life: typical 6 hrs, continuous testing. The IRM-5000P can be used while charging. |
PRINTER |
Built-in 2.5-inch wide thermal printer can print test results in both tabulated and graphic formats |
COMPUTER INTERFACE |
One RS-232C port |
SAFETY |
UL Certified (UL 61010A-1), CAN/CSA Certified (C22.2 No. 1010.1-92) |
ENVIRONMENT |
Operating: -10˚ to 50˚ C (15˚ to +122˚ F); Storage: -30˚ C to 70˚ C (-22˚ to +158˚ F) |
CABLES |
One 6-foot cable set, one 15-foot cable set, ground cable, power cord, cable bag |
OPTIONS |
Transportation Case |
WARRANTY |
One year on parts and labor |
Specifications are valid at nominal voltage and ambient temperature of +25˚C (+77˚F). Last updated on February 6, 2009 |